I-Test

Compare

Overview

When the Gas-Pro or T4 are inserted inserted into I-Test, no button presses are required and I-Test starts testing automatically. A bump test takes less than 1 minute. The test carries out verification that the sensors are responding to a known value of gas, and filters are clear and good for use. It will also ensure the audible and visual alarms are working, giving the user full confidence that they are compliant for site use. 

I-Test is easy to use with no requirements to turn on and off the gas: the design of the demand flow regulator coupled with the I-Test will allow the automatic pull of the correct required amount and concentration of gas. I-Test can also recognise cylinders that are empty or those that have expired allowing full compliance.

I-Test stores all bump and calibration results and this can then be interrogated using the I-Test Manager software. This powerful software allows you to carry out exception reporting to see who has and hasn’t carried out their bump or calibrations. I-Test Manager also makes it very easy to print both bump and calibration certificates and any tailored reports you have generated.

Flexible

  • Mains powered, or battery back-up – Use  I-Test on site, or in the field
  • Entirely self-contained – No extra computers or equipment required
  • Ability to update Gas-Pro and T4 configurations – For example, update the bump test interval or alarm levels of your fleet with ease
  • Data management – Gas-Pro and T4 store the gas levels they experience. Archive this data long term using I-Test.
  • Powerful support software – Makes configuring I-Test extremely easy.

Reliable

  • I-Test recognises empty or expired cylinders – Allows for planned refills and no unit downtime
  • Testing starts automatically when Gas-Pro or T4 are inserted. Minimal user training required
  • Battery backup – Testing continues even if the power is out

Records

  • Everything is logged – See everything that‘s been tested and whether it passed/failed
  • Exception reporting – Precise reports flag up who has and hasn’t carried out their bump test or calibrations
  • Powerful reporting – Need to see which detectors have gone into alarm in the last month? I-Test Manager makes easy work of pulling this summary together 

Specifications

Size 110 x 250 x 320mm (4.3 x 9.8 x 12.6 ins)
Weight 2kg (70.5oz)
Alarms 9-12V @ 3A
I-Test backup battery 200 speedy bump tests with 15 second intervals bewteen tests
5 hours in standby with no bump test
Battery life approximately 500 cycles
Charge time 7 hours
RTC backup battery 3 year life
USB interface One standard USB type A socket for USB Flash Memory stick
One standard USB type B socket for connections to PC
Inlets/Outlets Single gas inlet (requires demand flow regulator)
Purge inlet (air)
Exhaust
Regulators Demand flow regulator required
Storage capacity Internal 2GB MicroSD card
Temperature specifications Operating temperature: -10°C to +50°C
Storage temperature: -25°C to +60°C
Charging temperature: 0°C to +40°C
Humidity 0% and 95% relative humidity (non-condensing) at +40°C
Ingress protection IP20

Accessories

Below is a small selection of the accessories available for the I-Test & I-Test Manager.

Cylinder holder

 

Cylinder holder (wall mount) for 34l to 110l cylinders

Demand flow regulator

 

Calibration gas

 

Downloads

Datasheet

Manual